    CE Covid 19 Self Swab Test Kit , Antigen Self Testing Kits

    Ask Lasest Price
    Brand Name : Vchek
    Price : negotiable
    Delivery Time : 10-15days
    Payment Terms : T/T, Western Union
    • Product Details
    • Company Profile

    CE Covid 19 Self Swab Test Kit , Antigen Self Testing Kits

    2019 nCoV Ag Saliva Rapid Test Card

    How to read the results
    Positive result: You can see a line next to the letter C and a line next to the letter T.

    Negative result: You can see a line next to the letter C, and nothing next to the letter T.

    In all other cases, your test is void. If no lines appear, or only a line next to the letter T, this means your test has not worked and is void. In this case, do a second test with a new kit.

    Is your self-test positive? That means that you probably have coronavirus.
    You should always contact your doctor in this case, or request an activation code for a PCR test through the contact centre on 02 214 19 19. A positive result in a self-test must be confirmed by a PCR test by your doctor or at a test centre. If a self-test is positive, you must immediately self-isolate. If the positive result is confirmed by your doctor or at a test centre, contact tracing will be started. You must stay indoors for at least 10 days.

    Is your self-test negative? That means that you are probably not contagious at present.
    You do not need to self-isolate. Be aware that a self-test is only ever a snapshot and is also not 100% reliable. If you do develop symptoms, you should always consult your doctor. Continue to follow the applicable rules carefully


    MATERIALS AND COMPONENTS


    Materials provided with the test


    IngredientsCassetteInstructions for useQuick Reference Instructions
    Specifications
    0674C4X00111NA
    0674C4X00221NA
    0674C4X005311
    0674C4X0101011
    0674C4X0202011
    0674C4X0252511

    Materials required but not provided

    • Timer

    STORAGE AND STABILITY

    • Store the test as packaged between 2-30°C.
    • The Test stable until the expiration date printed on the outer packing, the product will be expired after 24 months.
    • Do not use beyond the expiration date.
    • Do not freeze any contents of the test
    • The test must remain in the sealed pouch until use.

    TEST PROCEDURE

    Before test, please read the instructions carefully.

    • Take the cassette to equilibrate to room temperature.
    • Unpack the aluminum foil bag, take out the cassette.
    • Insert the absorber end of the cassette into mouth. Make sure cassette is horizontal statement.

    CE Covid 19 Self Swab Test Kit , Antigen Self Testing Kits

    • Actively swab the inside of the mouth and tongue to collect oral fluid.
    • Remove the absorber end from the mouth when the purple color move across the result window in the center of the test device.
    • Wait for 10 minutes and read the results.

    NOTE:

    *When sampling, gently hold it in mouth and let saliva naturally adsorb on the sponge.

    *Do not eat, drink, or smoke prior to the test for at least 30 Minutes.

    *Any saliva specimen is appropriate for testing but the saliva specimen collected in the morning, before mouth rinsed, eating or drinking, is recommended


    INTERPRETATION OF TEST RESULTS

    This product can only perform qualitative analysis on the detection object.

    Positive Result:

    If both C and T lines are visible within 10 minutes, the test result is positive and valid.

    Negative Result:

    If test area (T line) has no color and the control area displays a colored line, the result is negative and valid

    Invalid Result:

    The test result is invalid if a colored line does not form in the control region. The sample must be re-tested, using a new test.


    INTERNAL CONTROL

    The test contains a built-in internal control in the test card. A color band appearing in the control region (C) is designed as an internal control. The appearance of the control line confirms that sufficient flow has occurred, and that the test card is working normally. If the control line does not appear within 10 minutes, it is considered an error in the test result and it is recommended to test again with the same sample and a new device.


    LIMITATIONS

    • The result of the test should not be taken as a confirmed diagnosis, for clinical reference only. Judgement should be made along with RT-PCR results, clinical symptoms, epidemiological information, and further clinical data.
    • The Test performance depends on the amount of virus (antigen) in the sample and may or may not correlate with viral culture results performed on the same sample.
    • The test must be equilibrated to room temperature (18℃~26℃) before used, otherwise the results may be incorrect
    • A negative test result may occur if the level of antigen in a sample is below the detection limit of the test.
    • Failure to follow the Test Procedure may adversely affect test performance and/or invalidate the test result.
    • React less than 10 minutes may lead a false negative result; React more than 10 minutes may lead a false positive result.
    • Positive test results do not rule out co-infections with other pathogens.
    • Negative test results are not intended to rule in other viral or bacterial infections.
    • Negative results should be treated as presumptive and confirmed with a molecular assay.
    • Clinical performance was evaluated with fresh samples.
    • Users should test specimens as quickly as possible after specimen collection.

    PERFORMANCE CHARACTERISTIC

    1. Clinical Verification

    The performance of Test was established with 232 sample collected from symptomatic patients, who with symptoms onset within 7 days.


    2019-nCoV Saliva Ag EASY TESTComparative RT-PCR Test Result
    (Immunochromatography)
    Positive (+)Negative (-)Total
    Detected Positive1081109
    Detected Negative7116123
    Total115117232
    Sensitivity93.91%, 95% CI (87.97,97.02)
    Specificity99.15%, 95% CI (95.32, 99.85)
    Accuracy96.55%, 95% CI (93.34, 98.24)

    Positive results broken down by CT value:


    2019-nCoV Saliva Ag EASY TESTComparative RT-PCR Method
    (Immunochromatography)(Positive by Ct Value)
    Positive (Ct<=25)Positive (25<Ct)
    Detected Positive6939
    Total7045
    Positive agreement98.57%86.67%

    2. Limit of Detection

    The experimental results show that for the virus culture concentration above 100 TCID50/mL, the positive rate of detection is greater than or equal to 95%. For the virus culture concentration of 50 TCID50/mL and below, the positive rate of detection is lower than 95%. So, the limit of detection of the Test Card is 100 TCID50/mL.

    3. Cross-reactivity

    Cross-reactivity of the test card was evaluated. The results showed no cross reactivity with the following specimen.


    4. Interference Substances

    The test results do not be interfered with the substance at the following concentration:


    5. Precision

    1. Test 10 replicates of negative and positive by using the reference materials of enterprises. The negative agreement and the positive agreement were 100%.

    2. Test three different lots kits including positive and negative reference materials of enterprises. The negative results and the positive results were 100%

    6. Hook Effect

    The Test Card was tested up to 1.6 × 105 TCID50/ml of heat-inactivated 2019-nCoV strain and no high-dose effect was observed.


    PRECAUTIONS

    • For in vitro diagnostic use.
    • Use appropriate precautions in the collection, handling, storage, and disposal of patient samples and used test contents.
    • Use of Nitrile, Latex (or equivalent) gloves is recommended when handling patient samples.
    • Do not reuse the used Test Card or saliva swab.
    • Should never open the foil pouch of the Test Card exposing it to the ambient environment until the Test Card is ready for immediate use.
    • Discard and do not use any damaged or dropped Test Card or material.
    • Inadequate or inappropriate sample collection, storage, and transport may yield false test results.
    • Sample collection and handling procedures require specific training and guidance.
    • To obtain accurate results, do not use visually bloody or overly viscous samples.
    • To obtain accurate results, an opened and exposed Test Card should not be used.
    • Testing should be performed in an area with adequate ventilation.
    • Wear suitable protective clothing, gloves, and eye/face protection when handling the contents of this test.
    • Wash hands thoroughly after handling.

    KEY TO SYMBOLS USED

    CE Covid 19 Self Swab Test Kit , Antigen Self Testing KitsMaterials IncludedCE Covid 19 Self Swab Test Kit , Antigen Self Testing KitsTest Cassette
    CE Covid 19 Self Swab Test Kit , Antigen Self Testing KitsInstructions for UseDate of
    Manufacturer
    Consult InstructionsDo Not Reuse
    For Use
    Store at 2°C~30°CCE Covid 19 Self Swab Test Kit , Antigen Self Testing KitsCatalogue Number
    CE Covid 19 Self Swab Test Kit , Antigen Self Testing KitsExpiration DateKeep away from Sunlight
    ManufacturerCE Covid 19 Self Swab Test Kit , Antigen Self Testing KitsTests per Kit
    CE Covid 19 Self Swab Test Kit , Antigen Self Testing KitsLot NumberCE Covid 19 Self Swab Test Kit , Antigen Self Testing KitsIn Vitro Diagnostic Medical Device
    Keep Dry
    Guangzhou Decheng Biotechnology Co., LTD
    Room 218, Building 2, No.68, Nanxiang Road, Science City, Huangpu District, 510000, Guangzhou P.R.China
    TEL:+86-020-82557192
    sales@dochekbio.com
    www.dochekbio.com
    CMC Medical Devices & Drugs S.L.
    C/ Horacio Lengo Nº 18, CP 29006,
    Málaga, Spain